APS TechNotes See all articles


ASTM Type I II and III Water Specifications 2010




Everything you need to make lab water
New Lab Water Systems
Replacement filters for all brands of existing lab water system

Type IType IIType IIIType IV
Electrical conductivity, max, S/cm0.0561.00.255.0
Electrical resistivity, min181.04.00.2
pH*** 5.0 to 8.0
Total organic carbon (TOC), max, g/L5050200no limit
Sodium, max, g/L1510 50
Chlorides, max, g/L1510 50
Total silica, max, g/L3 3500no limit

Bacterial contamination

When bacteria levels need to be controlled, reagent grade types should be further classified as follows:

Type A Type BType C
Maximum heterotrophic bacteria count10/1000 mL10/100 mL100/10 mL
Endotoxin, EU/mlB 0.030.25not applicable


*pH is not a significant indicator when comparing ionic contamination.

Also See :
  • Measuring pH of pure water
  • NCCLS and CAP Water Specs



    ATSM Specs







  • We Offer Over 22,000 Products
    ASTM Type I II and III Water Specifications 2010
    Purchase Orders Accepted
    (subject to approval)
    Credit Application

    Rated "A+" with the BBB
    24 years, Zero Complaints

    Got a Millipore Water System?
    For Huge Savings
    Click Here
    Best ...
    Independent Reps Welcome

    New Feature
    Lab Water Filter Database

    Millipore Filter Quick List
    Barnstead Filter Quick List

    Barnstead Micropure - ASTM Type I II and III Water Specifications 2010



    ... No Description... ... List All Products
    Google Apps

    APS Water Services Corp. 7320 Valjean Ave, Van Nuys, CA. 91406
    800-460-9011 Toll Free | 818-786-0600 Local/International | 818-786-2347 Fax | Public@apswater.com



    ASTM Type I II and III Water Specifications 2010
    Images may or may not be of actual product.

    * Flat Rate UPS shipping when able to ship via UPS and is in the USA excluding Hawaii and Alaska.
    Larger Items may not be able to ship via UPS, in that case freight charges will be quoted seperately.

    7/7/2015 at 8:25:59 AM
    Terms and conditions
    Credit Application

    Privacy Policy

    Tags :



    ASTM Type I II and III Water Specifications 2010