Im looking at RO10L1 I assume it can deliver close to 18 M ohm water Can you confirm
Answer
No, that would not produce 18 megohm. It is designed as pretreatment to a lab polishing system or to use for low quality applications such as glassware washing etc. In order to make 18 megohm water you need recirculation and instrumentation.
A good combination is the RO75L1 RO/DI system feeding a Barnstead Micropure System