Operating tips for a Barnstead Diamond Lab Water System
Operating tips for a Barnstead Diamond Lab Water System
Date : 3/26/2014
From :
Christi Richmond
Question
We use a Barnstead Nanopure Diamond UV system to produce HPLC grade water for our laboratory. Is there a need to flush the unit prior to dispensing water? Is there any value added in flushing the unit for two minutes prior to using the dispensed water? We use the unit in recirculating mode and wait until the resistivity is greater than 18.0 mOhm-cm before use.
Answer
There is no need to do a flush. Flushing would be more for removing built up endotoxin in an Ultrafiltration membrane (UV/UF systems).
If your measurement requirement is "ultra low", you will want to throw away the first 500 ml's of water you draw off because the bacteria filter is not recirculated and can build up some bacteria and the related TOC's from their waste.
Putting the unit in recirculation mode will assure the highest purity when you need it. Each time the water recirculated any TOC and ions are removed. The water is essentially "Polished".
For replacement filters and operating manuals you can click here :
This high capacity four housing lab water system is designed to provide up to 3 l.p.m. of ultrapure water exceeding ASTM and CAP Type I, II or III specifications.
Code F, 2.5 inch X 10 inch. Coconut shell carbon block with 10 micron outer wrap. Special Prep 1, for Chlorine removal with low extractable applications - 1/pk
Filter Housing
IWT 3C0600002 Style High Capacity Filter - Cartridge Universal II
This High Capacity Mixed Bed Deionization Filter is used for the reduction of dissolved minerals. with Color Indicator - Comparable to IWT and US FILTER 3C0600002 and HC200 or HC-200
Great for Lab Water Pretreatment
or Great Drinking Water
Images are representative of the products. Images may or may not be of the actual product. If it is important e-mail us for an actual image if available.
* Flat Rate UPS shipping when able to ship via UPS and is in the USA excluding Hawaii and Alaska. Larger Items may not be able to ship via UPS, in that case freight charges will be quoted seperately. International shipping will be quoted after the order is placed. You will have the opportunity to cancel before we finalize your order.