Nanopure Diamond Analytical that has an error message OVERPRESSURE ERROR
Shopping Cart Items: 0   Sub-Total : US$0.00
Welcome, Guest ( Login )
Celebrating 35 years providing high quality products and advice.
Our Local Time Is 4:35:27 AM.
Call us at 818-786-0600. We are here to help!
Laboratory Water Systems Laboratory Process Filters Laboratory Equipment Water Treatment

Nanopure Diamond Analytical that has an error message OVERPRESSURE ERROR


Date : 9/9/2014
From : Mastin

Question

I have a Nanopure Diamond Analytical that has an error message OVERPRESSURE ERROR In the manual it states that the problem is the 2u filter that needs to be replaced I did that and still have the problem Any Ideas Amy


Answer

Remove the filter, open the outlet valve to bleed off the pressure. Make sure the filter is full of pure water and then re-install it.

That may clear the problem, if not you may have a sensor, electronic or a feed water pressure problem.

New filters cnn be found here

https://www.apswater.com/shopdisplayproducts.asp?id=111&cat=Filters_for_Barnstead_Nanopure_Diamond_Systems

The operating manual can be found here :

https://www.apswater.com/article.asp?id=145&title=Manuals+for+Barnstead+Brand+Water+Systems