E-Pure and Nanopure II low resistivity in circulation but not draw off
Date : 11/24/2014
From :
Art
Question
Couple questions My E-pure system reads about 4 megaohms when circulating but goes up to 18.3 after dispensing for about 1 minute What would case that Also is there a way to filter dnase rnase free and low endotoxin levels using the Epure systems That is are there filters that would generate that purity of water
Answer
What you are describing is the failure of the water to recirculate. That can be caused by a few different things.
1. Bad pump 2. blocked return check valve 3. Bad of high feed pressure regulator
In order to make RNASE and DNASE Free water you must have ultrafiltration. You can not add that to an e-pure system. You would have to change to one of the UF models found here: